Set Measurement Speed; Nplcycles <N - Keithley 2500 User Manual

Photodiode meter
Hide thumbs Also See for 2500:
Table of Contents

Advertisement

17-56
SCPI Command Reference

Set measurement speed

NPLCycles <n>
[:SENSe[1]]:CURRent[:DC]:NPLCycles <n>
:SENSe2:CURRent[:DC]:NPLCycles <n>
Parameters
Query
Description
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
<n> =
0.01 to 10
DEFault
MINimum
MAXimum
:NPLCycles?
:NPLCycles? DEFault
:NPLCycles? MINimum Query minimum PLC
:NPLCycles? MAXimum Query maximum PLC
This command is used to set the integration period (speed) for measure-
ments. NPLC (Number of Power Line Cycles) expresses the integration
period by basing it on the power line frequency. For example, for a PLC
of 1, the integration period would be 1/60 (for 60Hz line power) which
is 16.67 msec.
Note that this is a global command. Thus, if you set the speed for chan-
nel 1 measurements to 10 PLC, then channel 2 will also set to 10 PLC.
Models 2500 and 2502 User's Manual
Set channel 1 speed (PLC)
Set channel 2 speed (PLC)
Power-line cycles per integration
1
0.01
10
Query programmed PLC value
Query *RST default PLC