Instrument Service Position
Refer to Chapter 4 for information on removing the spectrum analyzer
cover assembly and accessing all internal assemblies.
Table 2-1
Related Adjustments
Assembly
Perform the following related
Changed
adjustments in the order listed
or
Repaired
A1A1
No related adjustment
keyboard
A1A2 RPG
No related adjustment
A2
controller
If EEROM from old A2 controller could not be
used in new A2 or if EEROM must be replaced,
also perform the following adjustments:
LOMA adjustments
External mixer amplitude adjustment or
3rd amp/2nd IF align
Front end cal
A3 interface
Front end cal
A4 log
amp/cal osc
Demodulator adjustment
IF amplitude adjustment
DC log amplifier adjustment
A5 IF
IF bandpass adjustment
IF amplitude adjustment
A6 power
No related adjustment
supply
*If any automated adjustment is required, you must first perform automated
adjustment
"1. Initial Information" on page 63
Chapter 2
†
†
†
†
†
.
Adjustment/Diagnostic Software
Introduction
Adjustment Number
Manual
Automated*
4
11
5
7
7
9
2
9
3
10
1
8
2
9
51