141 BAD ARG Argument can only be used with FDIAG command.
142 BAD ARG Query expected for FDIAG command.
143 NO PRESL No preselector hardware to use command with.
144 COUPL?? Invalid COUPLING argument, expected AC or DC.

ADC Errors (200 to 299)

These errors are directly related to the ADC/interface
section. Suspect a faulty A2 controller, A3 interface
assembly or, in 8560 E-series analyzers with
Option 007, the A16 fast ADC (FADC) assembly.
Errors 202 through 207 apply only to EC-series
analyzers and E-series analyzers with fast ADC
(Option 007).
ADC driver/ADC hardware/firmware interaction; check
200 SYSTEM
for other errors.
ADC controller/ADC hardware/firmware interaction;
201 SYSTEM
check for other errors.
202 FADC CAL Binary search failed during FADC linear offset
calibration.
This error applies only to EC-series analyzers and
E-series analyzers with fast ADC (Option 007).
203 FADC CAL Binary search failed during FADC log offset
calibration.
This error applies only to EC-series analyzers and
E-series analyzers with fast ADC (Option 007).
204 FADC CAL Binary search failed during FADC log expand offset
calibration.
This error applies only to EC-series analyzers and
E-series analyzers with fast ADC (Option 007).
205 FADC CAL Slope derivation failed during FADC linear offset
calibration.
This error applies only to EC-series analyzers and
E-series analyzers with fast ADC (Option 007).
206 FADC CAL Slope derivation failed during FADC log offset
calibration.
This error applies only to EC-series analyzers and
E-series analyzers with fast ADC (Option 007).
Chapter 7
General Troubleshooting
Error Messages
329