11.Place the jumper on A14J23 in the NORM position (pins 1 and 2
jumpered).
12.Disconnect the SMA cable from the 1ST LO OUTPUT jack and
reconnect the 50Ω termination on the 1ST LO OUTPUT.
YTO FM Coil Adjustments
13.On the spectrum analyzer, press
follows:
Center frequency ..................................................... 300 MHz
Span ........................................................................... 20 MHz
14.Adjust A14R76 FM SPAN until the 300 MHz CAL OUTPUT
SIGNAL is aligned with the center vertical graticule line.
Chapter 3
Manual Adjustment Procedures
and set the controls as
PRESET
5. YTO Adjustment
133