Assembly Replacement

A11 YTO

A11 YTO

Removal
1. Carefully disconnect W43 semirigid cable from A9J2 and A12J3 and
move it out of the way.
2. Remove W56/FL2/W57 (as a unit) and disconnect W38 at the A11
assembly.
3. Remove four screws (1) securing A11 to the right-side frame. See
Figure 4-23, "A11 Mounting Screws at Right Side Frame."
the screws while holding onto A11.
4. Disconnect W10 ribbon cable from the A11 YTO assembly.
Replacement
1. Connect W10 ribbon cable to the A11 YTO assembly.
2. Orient the A11 assembly in the spectrum analyzer so its four
mounting holes line up with the holes in the right side frame and the
output connector is lined up with W38.
3. Secure the A11 assembly to the right side frame using four screws.
4. Connect W38 to A11.
5. Install W56/FL2/W57 between A12J1 and A13J1.
6. Reconnect W43 semirigid cable to A9J2 and A12J3 and torque all RF
cable connections to 113 Ncm (10 in-lb).