32.If the power is less than −6.5 dBm, suspect W34, A7 LOMA, or A11
33.Move jumper A14J23 to the NORM position.
Table 11-3
Sampling Oscillator Test Frequencies
8340A CW
Frequency
(GHz)
6.067000
6.087000
6.110200
6.141000
4.710000
6.174000
6.193000
6.213000
6.066000
6.069000
6.247000
6.107000
3.912000
6.289000
6.321000
6.333000
Chapter 11
YTO.
8564/5EC Center
Frequency
(MHz)
2156.3
2176.3
2199.5
2230.3
799.3
2263.3
2282.3
2302.3
2155.3
2158.3
2336.3
2196.3
1.3
2378.3
2410.3
2422.3
Synthesizer Section
Confirming a Faulty Synthesizer Section
Offset PLL
Counter
Sampling
Reading
Oscillator Freq
Sampler IF
(MHz)
(MHz)
285.000
82.000
286.364
73.364
287.500
72.700
288.462
83.308
288.889
87.778
290.000
84.000
290.909
83.909
291.667
88.000
292.500
76.500
293.478
94.044
294.444
63.667
295.000
88.000
296.000
64.000
296.471
63.118
297.000
84.000
297.222
91.333
533