Agilent Technologies 8564EC Service Manual page 30

Spectrum analyzers
Hide thumbs Also See for 8564EC:
Table of Contents

Advertisement

Table 1-4
Instrument
Power sensor
Other Equipment
Controller
Oscilloscope
Amplifier
Power supply
Signature multimeter
Recommended Test Equipment
Critical Specifications for Equipment
Substitution
Frequency range: 50 MHz to 50 GHz
Maximum SWR:
1.15 (50 to 100 MHz)
1.10 (100 MHz to 2 GHz)
1.15 (2.0 to 12.4 GHz)
1.20 (12.4 to 18 GHz)
1.25 (18 to 26.5 GHz)
1.30 (26.5 to 40 GHz)
1.50 (40 to 50 GHz)
Required to run operation verification
software and adjustment/diagnostic software
(8564EC and 8565EC)
Bandwidth (3 dB): dc to 100 MHz
Two channels
Minimum vertical deflection factor: ≤5 mV/div
Minimum timebase setting: <100 ns
Digitizing display with time cursors
Delta-t cursor accuracy in 500 ns/Div: <0.1 µs
Frequency range:
8564EC, 2.0 to 8.0 GHz
8565EC, 2.0 to 8.0 GHz
Minimum output power (leveled)
2.0 to 8.0 GHz: +16 dBm
Output SWR (leveled): <1.7
Output voltage: ≥24 Vdc
Output voltage accuracy: <±0.2 V
Clock frequency >10 MHz
Time interval function
Chapter 1
General Information
Recommended Test Equipment
Recommended
Use
Model
8487A
9816A,
9836A/C,
310, 320
332, 360
54501A*
P,A,T
11975
6114A
5005A/B
P,V
V
P
A
T
39