RF Section

High Band Problems

High Band Problems
1. If the low band is functioning, but amplitudes of signals >26.8 GHz
are wrong, check the A12 SBTX.
2. If the low band and millimeter band signals are correct, but
microwave band signals are wrong, check the A10 RYTHM.
3. Perform the steps located in
Drivers" on page
4. Troubleshoot the signal path. Refer to the power levels listed in
Figure 12-11 on page
Diagram.
600
"Control Latch for Band-Switch
615.
631, RF Section Troubleshooting Block
Chapter 12