Manual Adjustment Procedures

9 . Demodulator Adjustment

9
. Demodulator Adjustment
Assembly Adjusted
A4 log amplifier/cal oscillator assembly
Related Performance Test
There is no related performance test for this adjustment.

Description

A 5 kHz peak-deviation FM signal is applied to the INPUT 50Ω . The
detected audio is monitored by an oscilloscope. FM DEMOD is adjusted
to peak the response displayed on the oscilloscope.
Figure 3-18
Demodulator Adjustment Setup