13.If there are no errors after cycling the spectrum analyzer power, the
EEROM is working properly, but the frequency-response correction
data might be invalid. Check the spectrum analyzer frequency
response.
14.Place the WR PROT/WR ENA jumper in the WR PROT position.
15.Fold the A2 and A3 assemblies into the spectrum analyzer as
described in
"Procedure 5. A2, A3, A4, and A5 Assemblies."
Chapter 4
Assembly Replacement
Procedure 12 . EEROM