Manual Adjustment Procedures
Introduction

Adjustable and Factory-Selected Components

Table 3-2 on page 96
designation and name. For each component, the table provides a
description and lists the adjustment number.
Refer to
Table 3-3 on page 98
components used in the instrument along with their functions.
Factory-selected components are identified with an asterisk on the
schematic diagrams.

Adjustment Tools

For adjustments requiring a nonmetallic tuning tool, use fiber tuning
tool, part number 8710-0033.
Two different tuning tools may be necessary for IF bandpass
adjustments, depending upon the type of tuning slug used in the
slug-tuned inductors. If the tuning slug requires a slotted tuning tool,
use part number 8710-1010. If the tuning slug requires a forked tuning
tool, use part number 8710-0772.
Never try to force an adjustment control. This is especially critical
when tuning variable capacitors or slug-tuned inductors. Required
service accessories, with part numbers, are listed under
Chapter 1 , "General Information."

Instrument Service Position

Refer to
Chapter 4
for information on removing the spectrum analyzer
cover assembly and accessing all internal assemblies.
92
lists the adjustable components by reference
for a complete list of factory-selected
Service Kit,
in
Chapter 3