All but two of the microcircuit entries required (M4 and M5) are
NOTE
voltages listed on the A7 LOMA. Although the label on A7 lists "Int B4,
Int B5, SBTX B4, and SBTX B5", the software requires that you enter
only "Int B4" (Int4) and "SBTX B4" (SBTX4).
M4 and M5 values are listed on the tower (A10/A12).
If the program is interrupted (power failure or the like) before DONE is
pressed, none of the previously entered values under
Initial Information will be saved. If this occurs, rerun Initial
Information from the start.
15.When the last value required has been entered, press DONE.
16.Any of the listed adjustments can be selected next
The 83650A must be in "Analyzer" mode. Check the 83650A user
NOTE
documentation for the correct setting of the rear panel switch.
For some adjustments, the user prompts are displayed on the spectrum
analyzer display (DUT), not the controller display.
Use the 10 MHz reference out of the 83650A wherever a 10 MHz
reference input to another instrument in the setup is required, except
for adjustments using the 3335A. (The spectrum analyzer provides the
10 MHz reference to the 3335A.)
The program calls out an 8566, but an 8562A, 8563A, or 8563E can be
used to make spectrum analyzer measurements. The 8566B or an
8560 E-Series spectrum analyzer is recommended for speed of
measurement.
Chapter 2
(Initial Information must always be run first).
Adjustment/Diagnostic Software
Using the Adjustment/Diagnostic Software
61