General Troubleshooting
Introduction

Service Cal Data Softkey Menus

The jumper on A2J12 is shipped from the factory in the WR PROT
(write protect) position (jumper on pins 2 and 3). When the jumper is
set to the WR ENA (write enable) position (jumper on pins 1 and 2), an
additional service cal data menu is displayed under
.
Figure 7-5 on
CAL
page 323
illustrates those areas of the service cal data menu that are
available.
322
Chapter 7