Table 1-4
Instrument
AM/FM
signal generator
Counters
Frequency
standard
Microwave
frequency counter
Universal
counter
Receivers
Spectrum analyzer
(for Option 002)
Spectrum analyzer
Measuring
receiver
Sensors
General Information
Recommended Test Equipment
Recommended Test Equipment
Critical Specifications for Equipment
Substitution
Frequency range: 1 MHz to 200 MHz
Frequency modulation mode
Modulation oscillator frequency: 1 kHz
FM peak deviation: 5 kHz
Output frequency: 10 MHz
−10
Accuracy: <1 × 10
Frequency range: 9 MHz to 26.5 GHz
Timebase accuracy (aging): <5 × 10
External frequency reference input
Modes: TI A→B, frequency count
Time interval measurement range: 100 ns to
120 s
Frequency count range: 400 Hz to 11 MHz
Frequency resolution: 1 mHz
Timebase accuracy (aging): <3 × 10
External 10 MHz reference input
Frequency range: 300 kHz to 7 GHz
Relative amplitude accuracy:
300 kHz to 2.7 GHz: <±1.8 dB
300 kHz to 7 GHz: <±4.0 dB
Absolute amplitude accuracy:
3.9 GHz to 6.9 GHz: <±2.7 dB
Frequency accuracy:
<±10 kHz at 7 GHz
Frequency range: 300 kHz to 7 GHz
Amplitude range: −70 dBm to +20 dBm
Compatible w/power sensors
dB relative mode
Resolution: 0.01 dB
Reference accuracy: <±1.2%
38
Recommended
Model
8640B
8642A
5061B
5343A*
−10
Option 001
/day
5334A/B
−7
/month
8566B*
8566B*
8902A*
Use
A
P,A
P,A,M,V
P
P,A,T
A,T
P,A,T,
M,V