508 AMPL 30K Unable to adjust amplitude of 30 kHz resolution
bandwidth. Insufficient gain during LC bandwidth
calibration.
509 AMPL .1M Unable to adjust amplitude of 100 kHz resolution
bandwidth. Insufficient gain during LC bandwidth
calibration.
510 AMPL .3M Unable to adjust amplitude of 300 kHz resolution
bandwidth. Insufficient gain during LC bandwidth
calibration.
511 AMPL 1M Unable to adjust amplitude of 1 MHz resolution
bandwidth. Insufficient gain during LC bandwidth
calibration.
512 RBW <300 Unable to adjust resolution bandwidths less than
300 Hz. Insufficient gain during crystal bandwidth
calibration.
513 RBW 300 Unable to adjust 300 Hz resolution bandwidth.
Insufficient gain during crystal bandwidth calibration.
Unable to adjust 1 kHz resolution bandwidth.
514 RBW 1K
Insufficient gain during crystal bandwidth calibration.
Unable to adjust 3 kHz resolution bandwidth.
515 RBW 3K
Insufficient gain during crystal bandwidth calibration.
516 RBW 10K Unable to adjust 10 kHz resolution bandwidth.
Insufficient gain during crystal bandwidth calibration.
517 RBW 100 Unable to adjust 100 Hz resolution bandwidth. Crystal
sweep problem.
518 RBW 300 Unable to adjust 300 Hz resolution bandwidth. Crystal
sweep problem.
Unable to adjust 1 kHz resolution bandwidth. Crystal
519 RBW 1K
sweep problem.
Unable to adjust 3 kHz resolution bandwidth. Crystal
520 RBW 3K
sweep problem.
521 RBW 10K Unable to adjust 10 kHz resolution bandwidth. Crystal
sweep problem.
522 RBW 10K Unable to adjust symmetry of 10 kHz resolution
bandwidth in first crystal pole.
523 RBW 10K Unable to adjust symmetry of 10 kHz resolution
bandwidth in second crystal pole.
524 RBW 10K Unable to adjust symmetry of 10 kHz resolution
bandwidth in third crystal pole.