Agilent Technologies 8564EC Service Manual page 187

Spectrum analyzers
Hide thumbs Also See for 8564EC:
Table of Contents

Advertisement

Assembly Replacement
Procedure 11 . Rear Frame/Rear Dress Panel
6
.Remove the three screws securing the power-supply shield to the
power supply, and remove the shield. See (1) in
184.
7
.Disconnect the fan and line-power cables from A6J3 and A6J101 on
the A6 power supply assembly.
8
.Remove the two flathead screws that secure the rear-panel battery
assembly, and remove the assembly. Remove the battery and
unsolder the two wires attached to the battery assembly.
9
.Use a 9/16-inch nut driver to remove the dress nuts holding the BNC
connectors to the rear frame. If necessary, drill out the nut driver to
fit over the BNC connectors, and cover it with heatshrink tubing or
tape to avoid scratching the dress panel.
10
.Remove four screws that secure the rear frame to the main deck. See
(1) in
Figure 4-28 on page
11
.Remove the six screws that secure the rear frame to the left and
right side frames.
12
.Remove the knurled nut that secures the earphone jack. Carefully
remove the jack using caution to avoid losing the lock washer located
on the inside of the rear-frame assembly. Replace the washer and
nut onto the jack for safekeeping.
13
.Remove the rear-frame assembly.
14
.To remove the rear dress panel, remove the two nuts located on the
inside of the rear frame near the display adjustment holes.
240
241.
Figure 4-2 on page
Chapter 4