Procedure
The spectrum analyzer should be allowed to warm up for at least
NOTE
30 minutes before performing this adjustment.
1. Place the spectrum analyzer in the service position shown in
2. Zero and calibrate the 8902A/8482A combination in log display
3. Connect the 8482A through an adapter directly to the spectrum
4. Adjust A15R561 CAL AMPTD for a −10.00 dBm reading on the
Chapter 3
3-15 on page
134. Prop the A14 frequency control board assembly in
the service position.
mode. Enter the power sensor 300 MHz cal factor into the 8902A.
analyzer CAL OUTPUT connector.
8902A display.
Manual Adjustment Procedures
6. Calibrator Amplitude Adjustment
Figure
135