Unable to adjust 1 kHz resolution bandwidth. ADC
431 RBW 1K
handshake.
Unable to adjust 3 kHz resolution bandwidth. ADC
432 RBW 3K
handshake.
433 RBW 10K Unable to adjust 10 kHz resolution bandwidth. ADC
handshake.
434 RBW 300 300 Hz resolution bandwidth amplitude low in first
crystal pole.
435 RBW 300 300 Hz resolution bandwidth amplitude low in second
crystal pole.
436 RBW 300 300 Hz resolution bandwidth amplitude low in third
crystal pole.
437 RBW 300 300 Hz resolution bandwidth amplitude low in fourth
crystal pole.
1 kHz resolution bandwidth amplitude low in first
438 RBW 1K
crystal pole.
1 kHz resolution bandwidth amplitude low in second
439 RBW 1K
crystal pole.
1 kHz resolution bandwidth amplitude low in third
440 RBW 1K
crystal pole.
1 kHz resolution bandwidth amplitude low in fourth
441 RBW 1K
crystal pole.
3 kHz resolution bandwidth amplitude low in first
442 RBW 3K
crystal pole.
3 kHz resolution bandwidth amplitude low in second
443 RBW 3K
crystal pole.
3 kHz resolution bandwidth amplitude low in third
444 RBW 3K
crystal pole.
3 kHz resolution bandwidth amplitude low in fourth
445 RBW 3K
crystal pole.
446 RBW 10K 10 kHz resolution bandwidth amplitude low in first
crystal pole.
447 RBW 10K 10 kHz resolution bandwidth amplitude low in second
crystal pole.
448 RBW 10K 10 kHz resolution bandwidth amplitude low in third
crystal pole.
449 RBW 10K 10 kHz resolution bandwidth amplitude low in fourth
crystal pole.
450 IF SYSTM IF hardware failure. Check other error messages.