Manual Adjustment Procedures
Introduction
Table 3-2
Adjustable Components
Reference
Adjustment
Designator
Name
A4C707
FM DEMOD
A4R445
LIMITER
PHASE
A4R531
LOG AMP
TOS
A4R544
LIN
FIDELITY
BOW
A4R826
CAL OSC
AMPTD
A5L300
LC CTR 1
A5L301
LC CTR 2
A5L700
LC CTR 3
A5L702
LC CTR 4
A5R343
15 DB ATT
96
Adjustment
Description
Number
16
Adjusts the FM demodulation for a peak response.
5
Adjusts Limiter Phase for peak response.
5
Minimizes error to Top of Screen.
5
Minimizes Linearity Fidelity error.
4
Sets calibration oscillator output power (nominally
−35 dBm). This power is injected into
during the AUTO IF ADJUST routines.
3
Adjusts center frequency of first stage of LC
bandwidth filter to 10.7 MHz.
3
Adjusts center frequency of second stage of LC
bandwidth filter to 10.7 MHz.
3
Adjusts center frequency of third stage of LC
bandwidth filter to 10.7 MHz.
3
Adjusts center frequency of fourth stage of LC
bandwidth filter to 10.7 MHz.
4
Adjusts the attenuation of the reference 15 dB
attenuator for 15 db between minimum and
maximum attenuation.