Figure 4-8
A2, A3, A4, and A5 Assembly Removal
Replacement
1. Place the spectrum analyzer on its right side on the work bench.
2. Attach the assembly being installed to the two chassis hinges with
3. Leave the assembly in the folded-out position and attach ribbon
4. Attach all cables to the assembly, as illustrated in
5. Locate the cable clip on the inside of the right side frame. Make sure
Chapter 4
two panhead screws.
cables W1 and W2.
page
204.
that the coaxial cables are routed properly on the clip as illustrated
in
Figure 4-12 on page 206
Assembly Replacement
Procedure 5. A2, A3, A4, and A5 Assemblies
Figure 4-9 on
203