Adjustment/Diagnostic Software
Getting Started
Spectrum Analyzer Warm-up
Warm the spectrum analyzer up for at least 30 minutes before
performing the first adjustment.
Table 2-2
Test Equipment Required for Automated Adjustments
Description
Controller*
Synthesizer/level generator
Synthesized sweeper
Spectrum analyzer
Power meter
4.2 GHz power sensor
50 GHz power sensor
Frequency counter
Digital multimeter
Power splitter
Test and Adjustment Module (TAM)
Cable, 50 GHz coax
Cable, BNC-to-SMB (2 required)
Cable, BNC-to-BNC (3 required)
Cable, 36 inch SMA
Cable, GPIB (8 required)
* 4 megabytes of free memory is required for the adjustment program.
the required 83650A (programming language set to "analyzer" mode).
§
56
The software menu shows the 8340 as the synthesized sweeper, but it drives
The software menu shows the 8566 as the spectrum analyzer, but it will drive the other three
spectrum analyzers listed. If an 8566 is used, it must be an 8566B, not 8566A.
Must have calibration factors down to 10 MHz; a standard 8487A is only calibrated down to 50 MHz.
Model Number
HP 9000 model 236 (HP 9836) or
HP 9000 model 310 or
HP 9000 model 320 or
HP 9000 model 350
3335A
83650A
8566B (preferred) or
8563E (2nd choice) or
8563A or
8562A
438A
8482A
§
8487A
5342A
3478A
11667C
85629B
8120-6164
85680-60093
10503A
8120-1578
10833A/B