Agilent Technologies 8564EC Service Manual page 404

Spectrum analyzers
Hide thumbs Also See for 8564EC:
Table of Contents

Advertisement

17.Tune the source 1 kHz above the fractional N frequency. The voltage
18.If the DVM reading does not change, the A14 frequency control
Figure 11-3
YTO Loop Test Setup
Check A15 RF
19.Disconnect W34 from A15U100J1 and disconnect W32 from
assembly (steps
19-25)
20.Connect a frequency counter to A15J101. Connect a high-frequency
21.Connect a BNC cable from the spectrum analyzer 10 MHz REF
22.Set the 8340A/B to the following settings:
23.Set the spectrum analyzer to the following settings:
Chapter 11
measured on the DVM should be approximately −12 Vdc.
assembly is defective. Reconnect W32 to A14J501. Replace the
jumper on A14J23 to the NORM position.
A15J101.
test cable from an 8340A/B synthesized sweeper to A15U100J1. See
Figure 11-4 on page
IN/OUT to the 8340A/B FREQUENCY STANDARD EXT input.
Frequency standard ........................................................ EXT
Power level ................................................................. −5 dBm
Span .............................................................................. 0 Hz
Trigger ................................................................... SINGLE
Confirming a Faulty Synthesizer Section
532.
Synthesizer Section
531