RF Section

A8 Low Band Mixer

A8 Low Band Mixer

1. Connect the spectrum analyzer CAL OUTPUT to the INPUT 50 Ω
connector.
2. Set the spectrum analyzer as follows:
Center frequency ..................................................... 300 MHz
Span ................................................................................. 0 Hz
Input attenuation .......................................................... 10 dB
3. If the spectrum analyzer serial number prefix is 3641A or greater,
make sure A8 is receiving the −5 V and −4 V supply voltages from
frequency control board assembly A14 via cable assembly W12.
4. Using another spectrum analyzer, check for approximately −21 dBm
(300 MHz) at the input of A8. (This level can easily be measured at
the output of FL1 by disconnecting W45 from FL1.)
5. If the level at the input of A8 is less than −25 dBm, suspect FL1
low-pass filter, A10 RYTHM, A12 SBTX, or A9 input attenuator.
Refer to power levels shown on
Section Troubleshooting Block Diagram.
6. Check for approximately −30 dBm (3.9107 GHz) at the output of A8.
(This level can easily be measured at the output of FL2 by
disconnecting W57 from FL2.)
7. If the level at the output of A8 is less than −35 dBm, suspect A8 low
band mixer or FL2 low-pass filter.
604
Figure 12-11 on page
631, RF
Chapter 12