Manual Adjustment Procedures
5. YTO Adjustment
6. Adjust A14R93 3.2 GHz for the appropriate frequency counter
reading of 3.200 GHz ±1 MHz.
7. On the spectrum analyzer, press
8. Adjust A14R42 6.01 GHz for a frequency counter reading of
6.010 GHz ±1 MHz.
9. On the spectrum analyzer, press
10.Repeat steps 6 through 9 until both of these interacting adjustments
meet their tolerances.
Figure 3-14
YTO Adjustment Locations
132
.