Using the Adjustment/Diagnostic Software

To Load the Program

The 8564E/8565E Adjustment/Diagnostic Software needs to be
NOTE
installed on either an HFS formatted hard disk system or an SRM
system. It cannot be used with only a flexible disk drive.
1. If necessary, load HP BASIC into the computer memory. See the
2. Create a new directory on your hard disk drive or SRM node. For
3. Download the 8564EC & 8565EC Adjustment/Diagnostics
4. Unzip the download file to the directory created in step 2.
5. Use the MSI command to change to the mass storage device where
6. Access the directory (created in step 2) using the MSI command,
7. Type LOAD "PS_EDITOR",1 and press
8. Press New Sensor.
58
documentation for your computer for more information about
loading HP BASIC.
example, 8564eadj.
software from www.agilent.com/find/8560_software.
the new directory (created in step 2) is located. For example,
MSI ":,700,0" for an HFS system or MSI ":,21.0" for an SRM
system.
then type MSI "MANAGER/CAL_DATA/SENSOR_DATA". Press
(or
).
ENTER
RETURN
(or
).
RETURN
ENTER