Adjustment/Diagnostic Software

6. Cal Out Adjustment

6
. Cal Out Adjustment

Assembly Adjusted

A15 RF assembly

Related Performance Test

Calibrator Amplitude and Frequency Accuracy

Procedure

1. Connect the 8482A power sensor (through adapter) to the front
panel CAL OUTPUT on the spectrum analyzer.
2. Carefully follow the instructions issued by the software program.
3. If the spectrum analyzer has Option 103, TCXO frequency reference,
the TCXO will also be adjusted.
72
Chapter 2

Advertisement

Table of Contents