Adjustment/Diagnostic Software
9 . IF Amplitude
9
. IF Amplitude
Assembly Adjusted
A4 log amp/cal oscillator assembly A5 IF assembly
Related Performance Tests
IF Gain Uncertainty Scale Fidelity
Procedure
Figure 2-6
IF Amplitude Adjustment Setup
1. Connect equipment as shown in
Figure 2-6 on page 75
and carefully
follow the instructions issued by the software program.
Chapter 2
75