Manual Adjustment Procedures
Introduction
Table 3-3
Factory Selected Components
Reference
Adjustment
Designator
Number
A5C204
3
A5C216
3
A5C326
3
A5C327
3
A5C505
3
A5C516
3
A5C717
3
A5C718
3
98
Basis of Selection
Selected to optimize center frequency of LC tank that loads the crystal.
Selected to optimize center frequency of LC tank that loads the crystal.
Selected to optimize LC pole center frequency.
Selected to optimize LC pole center frequency.
Selected to optimize center frequency of LC tank that loads the crystal.
Selected to optimize center frequency of LC tank that loads the crystal.
Selected to optimize LC pole center frequency.
Selected to optimize LC pole center frequency.
Chapter 3