Agilent Technologies 8564EC Service Manual page 81

Spectrum analyzers
Hide thumbs Also See for 8564EC:
Table of Contents

Advertisement

Table 3-1
Related Adjustments
Assembly
Perform the following related
Changed or
adjustments in the order listed
Repaired
A1A1 keyboard
No related adjustment
A1A2 RPG
No related adjustment
A2 controller
If EEROM from old A2 controller could not
be used in new A2 or if EEROM must be
replaced, also perform the following
adjustments:
LOMA adjustments
External mixer amplitude adjustment or
3rd amp/2nd IF align †
Front end cal
A3 interface
Front end cal
A4 log amp/cal
osc
Demodulator adjustment
IF amplitude adjustment
DC log amplifier adjustment
A5 IF
IF bandpass adjustment
IF amplitude adjustment
A6 power
No related adjustment
supply
*If any automated adjustment is required, you must first perform automated adjustment
Initial Information" on page
Either the manual or the automated adjustment may be performed.
Chapter 3
63.
Manual Adjustment Procedures
Introduction
Adjustment Number
Manual
Automated*
4
11
5
7
7
9
2
9
3
10
1
8
2
9
"1.
93

Advertisement

Table of Contents
loading

This manual is also suitable for:

8565ec

Table of Contents