Manual Adjustment Procedures

6. Calibrator Amplitude Adjustment

6

Assembly Adjusted

A15 RF assembly

Related Performance Test

Calibrator Amplitude and Frequency Accuracy
Description
The CAL OUTPUT amplitude is adjusted for −10.00 dBm measured
directly at the front panel CAL OUTPUT connector.
Figure 3-15
Calibrator Amplitude Adjustment Setup
Equipment
134
. Calibrator Amplitude Adjustment
Measuring receiver ........................................................... 8902A
Power sensor ...................................................................... 8482A
Adapters
Type N (f) to BNC (m) ................................................ 1250-1477