Adjustment/Diagnostic Software
Introduction
Table 2-1
Related Adjustments
Assembly
Perform the following related
Changed
adjustments in the order listed
or
Repaired
A7 LOMA
LOMA adjustments
Front end cal
performance test in the 8560 E-Series and
EC-Series Spectrum Analyzer Calibration
Guide. The adjustment must be performed if
the performance test fails).
A8 low band
Front end cal
mixer
A9 input
Front end cal
attenuator
performance test in the 8560 E-Series Spectrum
Analyzer Calibration Guide. The adjustment
must be performed if the performance test
fails).
A10/A12
Front end cal
RYTHM/SB
TX
A11 YTO
YTO adjustment
LO frequency
YTO FM coil
A13 2nd
Front end cal
converter
A14
frequency
control
YTO adjustment
LO frequency
YTO FM cal
Front end cal
*If any automated adjustment is required, you must first perform automated
adjustment
"1. Initial Information"
Either the manual or the automated adjustment may be performed.
52
(or perform the frequency response
(or perform the frequency response
Adjustment Number
Manual
Automated*
4
7
7
7
7
5
2
3
7
5
2
3
7