Agilent Technologies 8564EC Service Manual page 409

Spectrum analyzers
Hide thumbs Also See for 8564EC:
Table of Contents

Advertisement

Synthesizer Section
General PLL Troubleshooting
PLL
YTO PLL
Reference PLL
Sampler PLL
Fractional N PLL
Figure 11-6
Unlocked PLL
2. If the integrator output voltage changes in the manner described in
step 1, the problem is external to the PLL. For example, the
reference frequency could be faulty. If the integrator output voltage
appears incorrect, confirm that the pulses out of the phase detector
(2) are attempting to tune the oscillator in the correct direction.
3. If the phase detector output is bad, check the inputs to the detector
(3). One input should be higher in frequency than the other; this
should match the phase detector outputs.
4. Confirm proper power levels for the signals at the input to the "N"
dividers (4), the reference inputs (5 and 7), and the loop feedback
path (6).
536
Measurement Point
A14J23 pin 1 (YTO ERROR)
A15J502 pin 3 (LO3 ERR)
A15J200 pin 13 (OFL ERR)
A14TP13 (INTEGRATOR)
Chapter 11