Agilent Technologies 8564EC Service Manual page 43

Spectrum analyzers
Hide thumbs Also See for 8564EC:
Table of Contents

Advertisement

Table 2-1
Related Adjustments
Assembly
Perform the following related
Changed
adjustments in the order listed
or
Repaired
A15 RF
10 MHz reference adjustment (TCXO,
Option 103) or CAL OUT adjustment †
Calibrator amplitude adjustment or CAL OUT
adjustment †
External mixer bias adjustment
Sampling oscillator adjustment
External mixer amplitude adjustment or 3rd
amp/2nd IF align
Front end cal
A15U100
Sampling oscillator adjustment
sampler
A19 GPIB
No related adjustment
A21 OCXO
10 MHz reference adjustment (OCXO)
*If any automated adjustment is required, you must first perform automated
adjustment
"1. Initial Information"
Either the manual or the automated adjustment may be performed.
Perform either the manual "YTO adjustment" (
"YTO FM coil" adjustments (2 and 3)
Chapter 2
5
) or the automated "LO frequency" and
Adjustment/Diagnostic Software
Introduction
Adjustment Number
Manual
Automated*
8
6
6
6
10
4
11
11
5
7
4
11
7