Adjustment/Diagnostic Software
7. Front End Cal
7
. Front End Cal

Assembly Adjusted

A10/A12 RYTHM/SBTX (tower) A14 frequency control assembly
A15 RF assembly

Related Performance Tests

Displayed Average Noise Level Frequency Response

Procedure

1. Carefully follow the instructions issued by the software program.
This test requires long waiting periods between adjustments. Total time
NOTE
required to complete the entire Front End Cal is approximately two (2)
hours.
Make sure that the 10 MHz reference output from the 83650A is
connected to the rear panel 10 MHz IN/OUT on the spectrum analyzer
being adjusted.
The software program will tell you, "Waiting for instrument to warm
up." If the spectrum analyzer being adjusted has already been on for
more than 30 minutes, press CONTINUE on the controller.
Chapter 2
73