IF Section
Cal Oscillator (P/O A4 Assembly)
Figure 9-25
Failed Crystal Set Symptoms
Low-Pass Filter
Refer to function block AB of A4 Log Amplifier Schematic Diagram
(sheet 4 of 4) in the 8560 E-Series Spectrum Analyzer Component Level
Information.
1. Connect a DVM positive probe to A4J9 pin 4.
2. On the 8564EC or 8565EC spectrum analyzer, press
3. Press
4. Observe the DVM reading while IF ADJUST STATUS: AMPLITUDE is
5. If the DVM reading is outside the range in step 3 but inside the
478
FULL IF ADJUST
displayed messages IF ADJUST STATUS: 300 kHz RBW and
IF ADJUST STATUS: 3 kHz RBW. During this time period, the
voltage should be within a 2 to 10 Vdc range.
displayed. The reading should be within the 2 to 10 Vdc range.
range in step 4, suspect one of the reactive components in the filter.
. Observe the DVM reading between the
.
CAL
Chapter 9