IF Section
Troubleshooting Using the Diagnostic Software
6. Select the 10 kHz sweep width in the software menu. The display on
7. Select the 4 kHz sweep width in the software menu. The display on
8. Select the 2 kHz sweep width in the software menu. The display on
9. If the cal oscillator is not sweeping, check the output of the sweep
Table 9-1
Sweep Width Settings
Sweep
Width
20 kHz
10 kHz
4 kHz
2 kHz
432
the second spectrum analyzer should be similar to that shown in
Figure 9-4 on page
433.
the second spectrum analyzer should be similar to that shown in
Figure 9-5 on page
434.
the second spectrum analyzer should be similar to that shown in
Figure 9-6 on page
434.
generator circuit (A4U804 pin 8 of function block Z). A series of
negative-going parabolas should occur. Frequency and amplitude
vary, depending on the sweep width chosen.
RANGE, MA0, and MA1 values for the sweep widths.
Sweep
Time
5 ms
10 ms
30 ms
15 ms
Table 9-1
Res BW
RANGE
Adjusted
A4U105
Pin 6
10 kHz
+5 V
3 kHz
+5 V
1 kHz
+5 V
300 Hz
+5 V
lists the
MA1
MA0
A4U105
A4U105
Pin 2
Pin 5
0 V
0 V
0 V
+5 V
+5 V
0 V
+5 V
+5 V
Chapter 9