Assembly Replacement
Procedure 13 . A21 OCXO

Replacement

Ensure that the insulator is installed between the A21 OCXO and the
CAUTION
main deck. Failure to do so will result in damage to the instrument by
shorting the power supply.
1. Connect W49, coax 82, to the OCXO and position the OCXO in the
2. Secure the OCXO to the spectrum analyzer main deck using three
3. Connect W50 to A15J306. Install tie wraps to hold W49 and W50
4. Fold the A14 and A15 assemblies into the spectrum analyzer as
5. Perform the rear frame assembly replacement procedure described
258
spectrum analyzer. Dress W50, orange cable, next to W49 through
the opening in the deck.
screws (1) and insulating washers. See
together.
described in
"Procedure
11
in
"Procedure
. Rear Frame/Rear Dress Panel."
Figure 4-34 on page
8
. A14 and A15 Assemblies."
257.