Sampling Oscillator Adjustment - Agilent Technologies 8564EC Service Manual

Spectrum analyzers
Hide thumbs Also See for 8564EC:
Table of Contents

Advertisement

Manual Adjustment Procedures

4. Sampling Oscillator Adjustment

4
Assembly Adjusted
A15 RF assembly
Related Performance Test
There is no related performance test for this adjustment procedure.
Description
The sampling oscillator tank circuit is adjusted for a tuning voltage of
5.05 V dc when the sampling oscillator is set to 297.222 MHz. The
voltage monitored is actually the tuning voltage divided by 4.05. The
setting is then checked at other frequencies for the full tuning range of
the sampling oscillator.
Figure 3-11
Sampler Adjustment Setup
Equipment
126
. Sampling Oscillator Adjustment
Digital voltmeter ............................................................... 3456A
DVM test leads ................................................................ 34118A
Chapter 3