Adjustment/Diagnostic Software
Using the Adjustment/Diagnostic Software
4. At the prompt ENTER STATION NUMBER WHERE THE ADDRESS WILL
5. Verify that the default GPIB addresses are the addresses actually
If you change any of the GPIB addresses, press RECORD ADR before
NOTE
continuing.
6. Press EXIT-NO RECORD.
7. Press Test Dut.
Be sure power to the 8564E/EC or 8565E/EC (DUT) is turned off before
CAUTION
connecting the 85629B TAM to the DUT. Failure to do so may cause
damage to the DUT or the TAM.
Initial Information must always be run first
NOTE
The WR PROT/WR ENA on the A2 controller must be in the WR PROT
(write protect) position.
8. Press Select, then CONTINUE.
9. Enter the power sensor "RP" numbers (last 4 digits of each serial
10.Enter employee number (or any number; an entry here is required to
11.Press DONE.
12.Select Initial Information.
13.Enter the serial numbers (up to 5 digits) for the "tower" (A10/A12
14.Enter the voltages printed on the microcircuits. Be sure to enter the
60
BE USED, type
, then
7
set on all of the instruments.
number) for the 8482A and 8487A (shows 8485A POWER SENSOR on
controller screen).
continue).
RYTHM/SBTX) and the LOMA (A7 LO Multiplier/Amplifier).
Entering these serial numbers is optional.
proper polarity, plus (+) or minus (−).
(or
).
RETURN
ENTER