Procedure 10. Bt1 Battery - Agilent Technologies 8564EC Service Manual

Spectrum analyzers
Hide thumbs Also See for 8564EC:
Table of Contents

Advertisement

Assembly Replacement
Procedure 10 . BT1 Battery
Procedure
Battery BT1 contains lithium polycarbon monofluoride. Do not
WARNING
incinerate or puncture this battery. Dispose of discharged
battery in a safe manner.
To avoid loss of the calibration constants stored on the A2 controller
CAUTION
assembly, connect the spectrum analyzer to the main power source and
turn on before removing the battery.
The battery used in this instrument is designed to last several years.
An output voltage of +3.0 V is maintained for most of its useful life.
Once this voltage drops to +2.6 V, its life and use are limited and the
output voltage will deteriorate quickly. When the instrument is turned
off, stored states and traces will only be retained for a short time and
may be lost. Refer to
10 , "Controller Section."
+2.6 V or less.
Removal/Replacement
1. Remove any option module attached to the rear panel.
2. Locate the battery assembly cover on the spectrum analyzer rear
3. Remove the old battery and replace it with the new one, ensuring
4. Measure the voltage across the new battery. Nominal new battery
5. Secure the battery assembly into the spectrum analyzer.
238
10
. BT1 Battery
"State- and Trace-Storage Problems,"
panel. Use a screwdriver to remove the two flathead screws securing
the cover to the spectrum analyzer.
proper polarity.
voltage is approximately +3.0 V. If this is not the case, check the
battery cable and A2 controller assembly.
The battery should be replaced if its voltage is
in
Chapter
Chapter 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

8565ec

Table of Contents