Calibrating for Increased Measurement Accuracy

Procedures for Error Correcting Your Measurements

Procedures for Error Correcting Your Measurements

This section has example procedures or information on the following topics:
• frequency response correction
• frequency response and isolation correction
• enhanced frequency response correction (with enhanced reflection error correction)
• one-port reflection correction
• full two-port correction (ES Analyzers Only)
• TRL*/LRM* correction (ES Analyzers Only)
• power meter measurement calibration procedure
NOTE
If the channels are uncoupled, you must make a correction for each channel.

Types of Error Correction

Several types of error correction are available that remove from one to twelve systematic
errors. The full 2-port correction effectively removes all twelve correctable systematic
errors. Some measurements do not require correction for all twelve errors.
explains each correction and its uses.
Table 6-2 Purpose and Use of Different Error Correction Procedures
Correction Procedure
Response
Response & Isolation
6-10
Corresponding
Measurement
Transmission or
reflection
measurement when
the highest accuracy
is not required.
Transmission of high
insertion loss devices
or reflection of high
return loss devices.
Not as accurate as
1-port or 2-port
correction.
Errors Corrected
Frequency response.
Frequency response
plus isolation in
transmission or
directivity in
reflection.
Table 6-2
Standard Devices
Thru for
transmission, open or
short for reflection.
Same as response
plus isolation
standard. (load)