Full Two-Port Error Correction (ES Analyzers Only)
• removes directivity errors of the test setup in forward and reverse directions
• removes source match errors of the test setup in forward and reverse directions
• removes load match errors of the test setup in forward and reverse directions
• removes isolation errors of the test setup in forward and reverse directions (optional)
• removes frequency response of the test setup in forward and reverse directions
This is the most accurate error-correction procedure. Since the analyzer takes
NOTE
both forward and reverse sweeps to update one measurement trace, this
procedure takes more time than the other correction procedures.
1. Set any measurement parameters that you want for the device measurement: power,
format, number of points, or IF bandwidth.
2. To access the measurement correction menus, press:
Cal
3. If your calibration kit is different than the kit specified under the
press:
CAL KIT
SELECT CAL KIT
If your type of calibration kit is not listed in the displayed menu, refer to
Calibration Kits" on page
4. To select the correction type, press:
CALIBRATE MENU
5. Connect a shielded open circuit to PORT 1.
NOTE
Include any adapters that you will have in the device measurement. That is,
connect the standard to the particular connector where you will connect your
DUT.
(select your type of kit)
7-57.
FULL 2-PORT
REFLECTION
Calibrating for Increased Measurement Accuracy
Full Two-Port Error Correction (ES Analyzers Only)
CAL KIT [ ]
RETURN
softkey,
"Modifying
6- 29