NOTE
You must use the floppy disk to store the following calibrations. Select the
floppy disk by pressing
3. Connect adapter A3 (same sex and connector type as the DUT) to adapter A2 on port 2
as shown in
Figure
Figure 6-15 Two-Port Cal Set 1
4. Perform a full 2-port calibration between ports 1 and 2 using calibration standards
appropriate for the connector type at port 1 (the connector type for adapter A1). Save
the calibration by selecting
5. Connect adapter A3 to adapter A1 on port 1 as shown in
Figure 6-16 Two-Port Cal Set 2
Save/Recall
6-15.
Save/Recall
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
SELECT DISK INTERNAL DISK
. Name the file "PORT1."
SAVE STATE
Figure
.
6-16.
6- 43