Saving Measurement Results

Instrument states combined with measurements results can only be saved to disk. Files
that contain data-only, and the various save options available under the
DEFINE DISK-SAVE
The analyzer stores data in arrays along the processing flow of numerical data, from IF
detection to display. These arrays are points in the flow path where data is accessible,
usually via GPIB. You can choose from three different arrays which vary in modification
flexibility when they are recalled.
• raw data
• data (raw data with error-correction applied) if correction is on; otherwise, raw data
• format (data processed to the display format)
If you choose to save the raw data array, you will have the most flexibility in modifying the
recalled measurement (including the ability to view all four S-parameters). This is because
the raw data array has the least amount of processing associated with it. Conversely, if you
choose to save the format array, your modification of the recalled measurement will be
limited by all the processes that are associated with that measurement result. However,
the format array is appropriate if you want to retrieve data traces that look like the
currently displayed data.
Define Save
Raw Data Array
Data Array
Format Array
You can also save data-only. This is saved to disk with default file names
DATA00D1 to DATA31D1 for channel 1
DATA00D2 to DATA31D2 for channel 2
DATA00D3 to DATA31D3 for channel 3
DATA00D4 to DATA31D4 for channel 4
However, these files are not instrument states and cannot be recalled.
key, are also only valid for disk saves.
Modification Flexibility During Recall
Most
Medium
Least
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
4- 39