Optimizing Measurement Results
Increasing Sweep Speed
To Set the Auto Sweep Time Mode
Auto sweep time mode is the default mode (the preset mode). This mode maintains the
fastest sweep speed possible for the current measurement settings.
• Press
Sweep Setup
To Widen the System Bandwidth
1. Press
IF BW
Avg
2. Increase the IF bandwidth to increase the sweep speed.
The specifications and characteristics chapter of the reference guide shows the relative
increase in sweep time as you decrease system bandwidth.
To Reduce the Averaging Factor
By reducing the averaging factor (number of sweeps) or switching averaging off, you can
increase the analyzer's measurement speed. The time needed to compute averages can also
slow the sweep time slightly, in narrow spans.
1. Press
AVG FACTOR
Avg
2. Enter an averaging factor that is less than the value displayed on the analyzer screen
and press
.
x1
3. If you want to switch off averaging, press
To Reduce the Number of Measurement Points
1. Press
NUMBER OF POINTS
Menu
2. Enter a number of points that is less than the value displayed on the analyzer screen
and press
.
x1
Refer to the "Specifications and Characteristics" chapter of the reference guide for
examples of how sweep time changes with the number of points.
To Set the Sweep Type
Different sweep speeds are associated with the following three types of non-power sweeps.
Choose the sweep type that is most appropriate for your application.
1. Press
Sweep Setup
2. Select the sweep type:
• Select
LIN FREQ
• Select
LIST FREQ
frequency points are of interest.
• Select
LOG FREQ
the lower part of the frequency span selected.
5-12
, to re-enter the auto mode.
SWEEP TIME
0
.
.
.
SWEEP TYPE MENU
for the fastest sweep for a given number of fixed points.
for the fastest sweep when specific non-linearly spaced
for the fastest sweep when the frequency points of interest are in
AVERAGING OFF
Avg
.
.