Figure 1-48 High Power Test Setup (Step 2a)
Figure 1-49 High Power Test Setup (Step 2b)
Selecting Power Ranges and Attenuator Settings
14.Select a power range that will not exceed the maximum estimated power level that will
force the DUT into compression. For example, if your booster amplifier has a gain of
+15 dB and the DUT will compress if supplied with +20 dBm, then you would adjust the
analyzer output power to not exceed +5 dBm by pressing
POWER RANGES
15.Adjust the analyzer power level to be below the maximum limit of the power range by
pressing
(or less)
0
RANGE 0 −15 TO +5
.
x1
Power
RANGE 0 −20 TO −5
(
Making Measurements
Measuring Amplifiers
PWR RANGE MAN
, 8722ES).
1- 61