Include any adapters that you will have in the device measurement. That is,
NOTE
connect the calibration standard to the particular connector where you will
connect your device under test.
Figure 6-8 Standard Connections for a One Port Reflection Error Correction
8. To measure the standard, when the displayed trace has settled, press:
If the calibration kit that you selected has a choice between male or female
NOTE
calibration standards, remember to select the sex that applies to the test port
and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the
calibration standard.
9. Disconnect the open, and connect a short circuit to the test port.
10.To measure the standard when the displayed trace has settled, press:
SHORT
The analyzer measures the short circuit and underlines the
11.Disconnect the short, and connect an impedance-matched load to the test port.
12.When the displayed trace has settled, press:
, select the type of load you are using, and then press
LOADS
the analyzer has finished measuring the load.
Notice that the
LOADS
13.To compute the error coefficients, press:
DONE: 1-PORT CAL
OPEN
softkey is now underlined.
Calibrating for Increased Measurement Accuracy
One-Port Reflection Error Correction
OPEN
softkey after it measures the
softkey.
SHORT
DONE: LOADS
when
6- 27