Using the Mixer Measurement Diagram
While the analyzer is still set to the IF frequency range, press:
INSTRUMENT MODE
System
LO MENU
FREQUENCY:CW
RETURN
RETURN
Note the RF frequency values on the diagram.
Press
Start
1.6

Perform a Power Meter Calibration Over the RF Range

1. Connect the equipment as shown in
Figure 2-18 Connections for Power Meter Calibration
2. Use the previous power meter settings.
3. Perform a one sweep power meter calibration over the RF frequency range at −5 dBm
(−10 dBm, 8722ES):
PWRMTR CAL
Cal
ONE SWEEP
−5
−10
(
x1
TAKE CAL SWEEP
FREQ OFFS MENU
1500
DOWN CONVERTER
G/n
Stop
2.5
G/n
Figure
.
, 8722ES)
x1
Making Mixer Measurements (Option 089 Only)
High Dynamic Range Swept RF/IF Conversion Loss
M/µ
RF > LO
.
2-18.
2- 23