Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal (ES Analyzers Only)
Figure 6-28 Two-Port Cal Set 2
7. Connect the ECal module between adapter A3 and adapter A2.
8. Press
ECal MENU
Cal
9. Press
FULL 2-PORT
module.
10.Save the results to disk. Name the file "PORT2."
11.Determine the electrical delay of adapter A3. If you have adapter specifications that
identify the electrical delay, you may use that information and continue with
the Adapter" on page
"Determine the Electrical Delay" procedure on
6-72
.
to perform the second 2-port error correction using the ECal
6-74. If you do not know the delay of adapter A3, perform the
page
6-73.
"Remove