Agilent Technologies 8719ES User Manual page 326

Network analyzer
Hide thumbs Also See for 8719ES:
Table of Contents

Advertisement

Calibrating for Increased Measurement Accuracy
Full Two-Port Error Correction (ES Analyzers Only)
17.To compute the error coefficients, press:
DONE 2-PORT CAL
The analyzer displays the corrected measurement trace. The analyzer also shows the
notation Cor at the left of the screen, indicating that error correction is on.
NOTE
You can save or store the measurement correction to use for later
measurements. Refer to
Measurement Results"
18.This completes the full two-port correction procedure. You can connect and measure
your device under test.
6-32
Chapter 4 , "Printing, Plotting, and Saving
for procedures.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8719et8720es8720et8722es8722et

Table of Contents